Multiparameter probe MPS-D8/Qualilog8

Robust, pluggable multi-parameter sensor from 1.4404 stainless steel to accommodate up to eight sensors (water level, temperature, conductivity, pH, redox potential, oxygen, turbidity, etc.)

Individually configurable:
Depending on the measuring task at hand, the MPS-D8 can be individually equipped with a maximum of eight water quality sensors and can detect up to 13 parameters.

Simple handling:
Special bayonet locks grant the users a simple and quick replacement of worn sensors. The calibration and configuration of the MPS-D8 is carried out using a special USB-RS485 interface cable together with ConfigPC operating software (Windows). This will automatically detect if a probe is spent and must be replaced.

Robust:
The housing of the MPS-D8 is made of high-quality 1.4404 stainless steel. The probe can thus be used even under adverse ambient conditions (e.g. landfill or other contaminated sites).

Variable:
The proven and robust SEBA plug system of the MPS-D8 allows both the connection to mobile handsets (SEBA KLL-Q-2 Type electric water level gauge (KLL), Checker-2) and stationary data collectors with data transmission (UniLogCom, LogCom/FlashCom) or without data transmission (Qualilog8).

Water Level 

Sensor Relative pressure probe
Measuring Principle Capacitive with ceramic measurement cell
Measuring Range 0-2 / 1 /4/ 5/ 20/ 40/ 100 / 200m Water column or according to specification
Measuring Accuracy ± 0.1% FS
Long Term Stability ± 0.1% / year

 

Water Temperature 

Sensor NTC30 with polynomial linearization
Measuring Principle NTC resistor
Measuring Range -5 - + 50 ° C
Measuring Accuracy 0.1 ° C
Pressure Range 0 - 50 bar

 

Conductivity 

Sensor 4-electrode measuring cell
Measuring Range 0-200mS / cm (total)
Measuring Accuracy ± 1μS / cm at measuring range 0-200μS / cm; +/- 0.5% with measuring range 0.2-200mS
Pressure Range 0 - 50 bar

 

pH 

Sensor Glass electrode
Measuring Range 0 - 14 pH
Measuring Accuracy ± 0.1pH
Pressure Range 0 - 20 bar

 

Redox

Sensor Glass electrode
Measuring Range -1200mV-1200mV
Measuring Accuracy ± 10mV
Pressure Range 0 - 20 bar

 

Oxygen

Sensor Clark electrode (amperometric)
Measuring Range 0-40mg
Measuring Accuracy ± 0.5% FS
Pressure Range 0 - 10 bar

 

Turbidity

Derived Parameters Suspended substances [mg / l] optically
Measuring Range 0-1000 NTU and 0-5000 NTU
Measuring Accuracy ± 1% of the respective end of measuring range
Pressure Range 0-10bar (with wiper) 0-20bar (without wiper)

 