WaterTechw2 Redox8000

The WaterTechw² Redox8000 Sensor has been designed to provide highly reliable Redox and Temperature measurements. The sensor uses a flat surfaced electrode which includes an extended reference path, these features combine to provide an extremely robust Redox measurement, suitable for use in surface water, waste water and drinking water applications.

Measurement of Redox or ORP is useful in many water, wastewater and industrial water processes, for example the control of disinfection processes such as chlorination in cooling towers. The electrode design is very similar to the equivalent pH electrode (WaterTechw² pH8000), the difference being that the pH glass is replaced with a Platinum band. The measurement of the oxidation-reduction potential (Redox or ORP) of an aqueous solution is a broad, non-specific indicator of the chemical activity of the solution. In drinking water processes Redox measurement is a highly effective tool for controlling chlorine and ozone treatment, it provides an excellent indicator of the ability of the disinfectant to remove contaminants from the water.

In wastewater processes Redox measurement has the ability to provide an indication of the condition of the anoxic and anaerobic zones within an activated sludge (ASP) plant’s aeration tanks, this is beyond what is possible with the far more commonly used Dissolved Oxygen measurements. The measurement can also be used as an indicator of the load entering a processing plant, giving an early warning of potential control problems.

Physical 

Dimensions  270 x ­Ø60 mm
Weight 1.2kg inc 10 metres of cable
Protection Class IP68
Wetted Materials Body: Black PVC, Nylon Cable Gland, Electrode: PPS (Ryton)
Seal Material Nitrile
Cable Type 4 core, 2 twisted pair, 5 mm O/D Polyurethane Coated
Cable Length 10 metres standard, 100 metres maximum
Temperature Rating 0 to +50°C
Pressure rating 2 bar
Depth Rating 20 m
Thread for Mounting Shaft Connection 1 1/2" BSPP Male


Interface to Monitor 

Power Supply 12 VDC from Monitor
Sensor Output ModTechw², RS485

Measurement Characteristics 

Principle Redox Electrode with Extended Reference Path
Temperature: Silicon Based Linear Thermistor
Applicable Units mV, °C
Range Redox: -1999 to +1999 mV
Temperature: 0 to 50°C
Accuracy Redox: +/- 2 mV
Temperature: +/- 0.5°C
Resolution Redox: 1 mV
Temperature: 0.1°C