TurbiTechw2 LS
The TurbiTechw² LS Sensor is designed to monitor the levels of suspended solids or turbidity typically found in final effluent from municipal and industrial wastewater treatment plants. It is also great at monitoring solids in the intake to both effluent and drinking water treatment processes.
Monitoring of these key parameters requires sensors that are reliable and accurate, the TurbiTechw² LS meets this requirement perfectly. The large optical surfaces and sample volume ensure that the information provided by the sensor is both representative of the process and tolerant of fouling. In addition, the sensor incorporates a self cleaning mechanism that ensures that the optical surfaces are kept clean at all times, the cleaning system is designed to avoid problems with ragging and does not smear the optical system.
The TurbiTechw² LS Sensor employs infrared light at 860 nm using the 90º Light Scatter principle in accordance with ISO7027 (2000). The cleaning process is automatically initiated by the 7300w² Monitor at a user determined frequency. The cleaning process takes only 90 seconds, which means that the sensor is available for 99.5% of the time with a 6 hourly cleaning cycle.
Physical
Dimensions | 470 x Ø80 mm |
---|---|
Weight | 2.2 kg (inc 10 metres of cable) |
Protection Class | IP68 |
Enclosure Material | Black Acetal Co-polymer |
Cable Entries | Integral Cable Gland |
Wetted Parts | Black Acetal, 316 Stainless Steel, Glass, Polyurethane, Nitrile |
Cable Type | 4 core, 2 Twisted Pair, 9 mm O/D Polyurethane Coated |
Cable Length | 10 metres standard, 100 metres maximum |
Service Requirement | Automatic Self Cleaning Seal Service every 3500 cleans (application dependent) |
Environmental Data
Operating Temperature | 0 to 50ºC |
---|
Interface to Monitor
Power Supply | 12 VDC from Monitor |
---|---|
Sensor Output | ModTechw², RS485 |
Range | 0-100 FTU |
---|---|
Accuracy | ±5% of measured value or 0.5 NTU whichever is greater |
Limit of Detection | 0.5 FTU |
Minimum Operating Range | 0 - 10 FTU |
Range | 0-500 FTU |
---|---|
Accuracy | ±5% of measured value or ±2.5 FTU whichever is greater |
Limit of Detection | 2.5 FTU |
Minimum Operating Range | 0 - 50 FTU |
Range | 0-1,000 FTU |
---|---|
Accuracy | ±5% of measured value or ±5 FTU whichever is greater |
Limit of Detection | 5 FTU |
Minimum Operating Range | 0 - 100 FTU |
Measurement Principle | 90° Light Scatter ISO 7027 Compliant |
|
---|---|---|
Wavelength | 860 nm Infrared | |
Pressure Rating (Depth) | 1Bar or 14.5 psi | |
Flow Rate | Not affected by flowrate, avoid deadspots and extreme turbulence | |
Units |
|