TurbiTechw² D-ISO Sensor
It’s an extremely sensitive sensor can detect changes in Turbidity of less than 0.01 FTU. It’s global standard of light scatter technique makes the sensor sensitive to a wider range of particle sizes.
The TurbiTechw² D-ISO Sensor is now SPAN (Suruhanjaya Perkhidmatan Air Negara) certified!
The sensor requires very little maintenance, with no tools or training required. Calibration can be carried out against either a wet Turbidity standard or using the dry calibration reference cell that is supplied with every sensor. The reference cell provides an extremely easy, repeatable method of checking the performance of the sensor, without the cost and inherent errors associated with chemical standards.
Physical
Dimensions | 300 x 185 x 128 mm (HxWxD) |
---|---|
Weight | 3.9 kg (inc 5 m of cable) |
Protection Class | IP65 |
Cable Entries | Integral cable gland |
Wetted Parts | Black acetal, Polypropylene, Nylon, chrome plated brass, Spectrosil 2000, nitrile |
Process Connections | Inlet and drain: ½" BSP with 12 mm hosetail Outlet: ¼" BSP with 5 mm hosetail |
Cable Type | 4 core, 2 twisted pair 5 mm outside diameter, polyurethane coated |
Cable Length | 5 metres as standard, other lengths can be supplied on request |
Interface to Monitor
Power Supply | 12 VDC from Monitor |
---|---|
Sensor Output | ModTechw², RS485 |
Measurement
Low Range | 0-30 NTU |
---|---|
Accuracy | ± 2% of reading or 0.02 NTU whichever is greater |
Limit of Detection | 0.02 NTU |
Minimum Operating Range | 0-3 NTU |
High Range | 0-200 NTU |
---|---|
Accuracy | ± 2% of reading or 0.1 NTU whichever is greater |
Limit of Detection | 0.5 NTU |
Minimum Operating Range | 0-20 NTU |
Measurement Principle | 90º Light Scatter, ISO 7027 Compliant |
---|---|
Pressure Rating (Max) | 2 bar |
Flow Rate | 0.5 L/min to 3 L/min |
Units | NTU, FTU, FNU |